Bernafon AG
Bern, 3018, Switzerland
NCT Number: NCT03086018
The aim of this study is to investigate the performance of the hearing aid under laboratory as well as everyday conditions. In addition, the performance data of the hearing aid are compared with a Bernafon hearing aid already available on the market.
The hearing aids are adapted according to the user requirements. The data collected in the study is intended to show that the benefit of the hearing aid being examined is identical or better than the current CE-marked hearing aid. A further aim is to improve the adaptation of the hearing device in such a way that the benefit for people with hearing impairments can be increased.
Looking for future studies?
Notify Me18 year and older
All sexes
Interventional
Not applicable
Bern, 3018, Switzerland
The reason for this study is to evaluate a new hearing aid. The goal is to evaluate the audiological performance and usability as well as features and functions in comparison to the hearing aid already CE marked. Furthermore, it is important to identify unexpected or unwanted behavior from the devices. This clinical investigation is a validation testing and is designed to evaluate the new hearing instrument system. As human subjects are involved, this validation test falls under the definition of a clinical investigation. The validation addresses the performance of the device with new functionality, and additionally addresses whether the end user can understand speech as well as with the current marketed device.
End users will be tested in the lab with speech tests and usability tests. They will also participate in field tests after which they will answer questionnaires concerning their experience with the instruments. Evaluating the overall performance of the hearing instrument is important to verify that the end user is satisfied with the device and that all user requirements are fulfilled. Many of the features available in the new device have already been validated and are used in devices currently on the market.
The purpose of this research is to show that the performance of the new device is as good as or better than the current device. Furthermore, speech should not be negatively affected with the addition of new features, the usability of the devices (that end users are able to handle the devices) should be as good as the current device, and there should be no artifacts or unwanted noises. Some measurements serve as quality control prior to product launch.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The participants will all be experienced hearing instrument device users and will be switched from their current (control) devices to the new test device which succeeds the Juna device.
Time frame: 2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
Speech test scores will be measured with speech reception thresholds (SRTs) of speech in noise in three conditions: unaided, aided with the current device, and aided with the new device. The score is measured by the the signal-to-noise ratio (SNR) at which 50% of words in a sentence list are correctly repeated. The speech is always presented at 65 dB and the background noise varies to maintain the 50% correct. A lower SNR score indicates a better score. The maximum score will be 15 dB and the minimum will be -5 dB.
Time frame: A 2 week time period during which they will wear the devices and answer questionnaires about the experience.
The subjective performance for the new device will be measured with the APHAB questionnaire (Abbreviated Hearing Aid Benefit Profile). The questionnaire is a list of 24 statement that the subjects must agree or disagree with using a scale of units from A to G (A being Always and G being Never). Each letter has a corresponding score used for the calculation: A receives 99, B receives 87, C receives 75, D receives 50, E receives 25, F recevies 12, and G receives 1.
An average for each subscale is calculated, and a global or overall score can be calculated by taking the mean of the three positive subscales (Ease of communication, Background noise, and Reverberation). For these three subscales a higher score indicates a better performance. The highest possible score is 99 and the lowest is 1. The fourth scale (Aversiveness) is a negative scale meaning that a higher score means a worse performance. The highest possible score is 99 and the lowest is 1.
Time frame: 2 hour period of speech testing during a lab appointment after wearing the devices for approximately 2 weeks.
Word recognition test scores will be measured with two signal-to-noise ratio (SNR) levels in percent correct in two conditions, with the new feature on and the new feature off.
The test is a speech test with the words stated in noise at either a 5 dB SNR or a 15 dB SNR. The subject is presented with 5 rhyming words and must choose which word they heard. It is a forced choice meaning that the test will not continue with the next word until they've made a selection. They cannot leave any blank and are not penalized for wrong answers. The score is a percentage of words correct with the maximum score being 100% and the minimum being 0%. A higher score means a better outcome.
Time frame: 2 hour period of usability testing with tasks and questionnaires.
The usability of the new device will be measured with a 3-point rating scale used by the tester to evaluate tasks performed by the end user as well as questionnaires and interviews.
The questionnaire asks subjects to perform tasks, and the clinician rates the ease with which they complete the tasks. 2 is the highest score meaning that they could complete without help, 1 means that they required some assistance, and 0 means that they could not complete the task. A higher score means a better outcome.
Time frame: A 2 hour period of interviews and questionnaires after having worn the devices for approximately 2 weeks and keeping a diary to make notes of device behavior.
Events will be measured by the reporting of unexpected sounds or behavior of the device through questionnaires, diaries, and interviews. The number of events reported including AEs related to the device will be used to measure the outcome. Less events is a better outcome.
Bernafon AG
Industry
A Comparative, Controlled, Clinical Investigation and Quality Control of a New Hearing Aid in Comparison to the Currently Marketed Device
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05219474
Ear Diseases, Hearing Disorders
New York, United States
View Trial DetailsNCT01805167
Ear Diseases, Hearing Disorders
Nice, France
View Trial DetailsNCT05070429
Aging, Ear Diseases
Hagerstown, Maryland, United States
View Trial DetailsNCT07501312
Ear Diseases, Hearing Disorders
Aurora, Illinois, United States
View Trial Details