NCT Number: NCT01469715
The purpose of this study is to assess the accuracy of the Becton Dickenson (BD) Technologies Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) in patients with Type 1 diabetes during low (goal glucose 55 mg/dL), normal (80-140 mg/dL) and high (>180mg/dL) glucose levels over a 24 hour period. This will be achieved by monitoring blood sugar levels when a regular dinner meal is given, when a liquid breakfast meal (BOOST Original containing 41 grams Carbohydrates (CHO), 4 grams fat, 10 grams protein) is given, when subcutaneous insulin is dosed to induce hypoglycemia to a goal of 55 mg/dL, and when a regular lunch meal is given. A Continuous Glucose Monitor (CGM) is an electronic device that measures and displays blood sugar (glucose) levels in the body throughout the day and night. The method being used to detect blood sugar in the investigational Glucose Binding Protein-Based Continuous Glucose Monitor (GBP CGM) is different than the method that is currently in use by commercially available models.
Some sensors cannot tell the difference between glucose (sugar) and other substances such as Tylenol, aspirin or citric acid etc. Because they cannot tell the difference, they may give false readings. The GBP CGM is made to only recognize glucose in the body rather than other substances (e.g., Tylenol, aspirin, citric acid, etc.). As a result, the investigators expect the new GBP CGM to be more accurate at detecting low blood sugar levels than the current devices.
Looking for future studies?
Notify Me21 year–65 year
All sexes
Interventional
Phase 2
Charlottesville, Virginia, 22908, United States
Development of accurate continuous glucose monitoring devices is critically important for the maintenance of strict glycemic control without increasing the risk for hypoglycemia and for the successful implementation of an artificial pancreas. Glucose binding protein-based (GBP) continuous glucose monitors offer several potential advantages over commercially available glucose oxidase based sensors. Unlike glucose oxidase based sensors, GBP sensors do not require establishment of an equilibrium reduction/oxidation reaction and are not as susceptible to interference from other compounds because they are engineered to recognize only glucose. As a result, GBP-based glucose sensors are exceptionally accurate at low blood glucose levels and have a faster warm-up time than glucose oxidase based systems. The principal idea of this proposal is to evaluate the accuracy and performance of a second generation glucose binding protein-based continuous glucose monitor (GBP CGM) developed by BD Technologies in subjects with type 1 diabetes, with emphasis on warm-up period after insertion, accuracy in hypoglycemic and hyperglycemic ranges, and lag time over a 24 hour period of sensor use. In order to accomplish this, the investigators will conduct a clinical trial consisting of 25.5-hour studies involving 15 subjects, each wearing one active and one mock (no GBP) device simultaneously during meals and induced hypoglycemia to observe a wide range of glucose values. The investigators envision that the glucose binding protein-based continuous glucose monitor will demonstrate minimal warm-up time after insertion, overall accuracy comparable to commercial glucose oxidase sensors with excellent performance in the hypoglycemic range and a nominal lag time to sensing a change in blood glucose.
Healthy volunteers accepted: No
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Visit 1: Screening visit to determine if subject qualifies for the study. Visit 2: Inpatient admission requiring a 25.5-hour hospital stay. Each subject will wear one active & one mock device simultaneously during hyperglycemic & hypoglycemic challenge conditions to observe a wide range of glucose values. Visit 3 & 4: Subjects will return to the research center approximately 24 & 48 hours after sensor removal, respectively, for evaluation of the postimplantation sensor site. Visit 5: Subjects will return to the research center approximately 28 days post inpatient admission. Blood samples for future testing of GBP and polyethylene Glycol neutralizing antibodies will be taken at Visit 1 & 5.
Other names: Glucose binding protein-based continuous glucose monitor
Time frame: 25.5 hours
ARD=100*(G_sensor-G_reference)/G_reference
Calculated for when patient's G_ref was Normal (70-180 mg/dl), Hyperglycemic (>180 mg/dl) and Hypoglycemic (<70 mg/dl)
The study data includes 208 paired sensor-YSI plasma glucose readings (G_reference) for each GBP CGM sensor (G_sensor) inserted for 24 hours during hyperglycemic and hypoglycemic challenge conditions. Data pairs will permit the detailed evaluation of sensor performance parameters, including static accuracy metrics such as median and mean absolute deviations and median and mean absolute relative deviation and Point CG-EGA, as well as dynamic parameters, such as warm-up time, trend accuracy (Rate CG-EGA), and sensor lag.
University of Virginia
Other
A Clinical Study of the Warm-Up, Accuracy, and Lag Performance of a Glucose Blinding Protein-Based Continuous Glucose Monitor (GBP CGM)
Acronym: SMA002
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07696039
Autoimmune Diseases, Diabetes Mellitus
Bergamo, Italy
View Trial DetailsNCT03263494
Autoimmune Diseases, Diabetes Mellitus
Los Angeles, California, United States
View Trial DetailsNCT07090824
Autoimmune Diseases, Diabetes Mellitus
Palo Alto, California, United States
View Trial DetailsNCT05211479
Adolescent, Anxiety Disorders
Yalova, Turkey (Türkiye)
View Trial Details