During right pulmonary vein ablation, the Circle Safe catheter is advanced through the venous system and positioned in the superior vena cava. Its distal electrodes provide diagnostic stimulation of the right phrenic nerve, while its proximal electrodes record the resulting diaphragmatic compound motor action potential (CMAP).
The recorded signal is transmitted to the Circle Safe System Console, which displays the CMAP waveform and automatically compares its amplitude with a baseline measurement obtained before ablation. This allows changes in diaphragmatic activity to be monitored throughout treatment of the right pulmonary veins.
The Circle Safe System is a diagnostic monitoring aid. It does not deliver ablation energy, modify the ablation procedure, or replace standard phrenic nerve monitoring methods used by the physician. Device function, signal stability, procedural use, and any device deficiencies or adverse events are documented during the study.