Clinical Research Facility
Birmingham, England, B15 2GW, United Kingdom
NCT Number: NCT07813091
This study aims to test the oxygen saturation component of the newly developed NeWTS monitoring system and compare readings from NeWTS to a traditional wired clinical monitor. It is important the investigators do this study in healthy adults as part of the development of this system, to make sure the oxygen saturation readings are accurate and before NeWTS is used in Neonatal Intensive Care (NICU).
Trial opening soon.
Get Notified18 year–55 year
All sexes
Interventional
Early Phase 1
Birmingham, England, B15 2GW, United Kingdom
The aim of the controlled desaturation study is to show that the NeWTS device can safely measure blood oxygen content in comparison to the "gold-standard" method of measuring blood oxygen levels (SaO2). To assess this, blood oxygen levels will be measured while the amount of oxygen participants breathe is gradually reduced.
Healthy adults will start breathing oxygen at the same levels as in room air (21% oxygen) and this will be gradually reduced. The measurements from the NeWTS device will be compared to the value taken directly from the blood via an arterial line (SaO2). The SaO2 is measured by obtaining multiple blood samples from an artery in the body.
As described in EN ISO 80601-2-61:2026 a minimum of 24 participants will be recruited with an average of 20 (±4), range 16-30 data points per participant.
The investigators will include a distribution of data pairs, with total pairs split across each desaturation decade between 70 % to 100 % saturations, with approximately 33 %, but no less than 25 %, in each decade.
90 % or greater of participants shall provide at least one useable data pair with an SaO2 below 85 %.
At least 75 % of participants shall also have provided at least one useable data pair with an SaO2 between 70 % to 80 %.
The number of data pairs for a given SaO2 decile (i.e. 70 % to 80 %, 80 % to 90 % and 90 % to 100 %) will be derived approximately equally from participants across skin pigment categories, ensuring that >25 % of data points come from light, medium and dark participants by MST and ITA.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
The primary objective for this clinical investigation is to validate the SpO2 performance of the NeWTS in comparison to "gold-standard" measurements of blood SaO2 by a calibrated and QC'ed ABL90 Flex plus blood gas analyser.
Time frame: From enrolment to end of study 3-4 hours
Validation of the SpO2 performance of the NeWTS in comparison to "gold standard" measurements of blood SaO2 by a calibrated and QC'ed ABL90 Flex plus blood gas analyser.
Contact information is provided by the study sponsor or research team.
Catherine Guy, BSc(Hons)
CONTACT
Kathryn Beardsall, MD
CONTACT
Kathryn Beardsall
Other
Controlled Oxygen Desaturation Study of the Neonatal Wireless Transmission System (NeWTS): NeWTS-DS
Acronym: NeWTS-DS
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07797556
Device Convenience, Device Performance
Houston, Texas, United States
View Trial DetailsNCT07527000
Device Performance, Healthy Adult Volunteers
Louisville, Colorado, United States
View Trial DetailsNCT07348289
Device Performance, Emergency Medicine
Park City, Utah, United States
View Trial DetailsNCT07284251
Device Performance
Malatya, Battalgazi, Turkey (Türkiye)
View Trial Details