NCT Number: NCT07749209

Agreement Among Three Methods of Intraoperative Core Temperature Monitoring: Nasopharyngeal Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Skin Sensor

Keeping a patient's body temperature normal during surgery is an important part of safe anesthetic care. When the body temperature drops below 36°C (called inadvertent perioperative hypothermia), patients face higher risks of bleeding, wound infection, heart problems, and a slower recovery. To prevent these complications, anesthesiologists monitor core body temperature continuously during surgery.

Three monitoring methods are commonly used in clinical practice: (1) a nasopharyngeal probe - a small soft sensor inserted through the nose and placed at the back of the throat, which provides continuous readings and is considered the gold standard; (2) an infrared tympanic thermometer (TaiDoc TD-1261) - a handheld device that measures ear temperature in about one second; and (3) the 3M SpotOn™ sensor - a non-invasive adhesive patch placed on the forehead that continuously estimates core temperature without requiring insertion into any body cavity.

Although all three methods are used interchangeably at Srinagarind Hospital, no study has directly compared all three simultaneously in the same patient to determine how well they agree with each other. If significant differences exist between methods, a clinician who switches from one device to another may unknowingly under-detect or over-estimate hypothermia.

This study will enroll 92 adult patients scheduled for elective surgery under general anesthesia at Srinagarind Hospital, Khon Kaen University. All three temperature sensors will be applied at the same time, and readings will be recorded simultaneously every 15 minutes throughout the operation. The main goal is to determine whether each device gives results close enough to the nasopharyngeal probe to be safely used as an alternative. The study is observational - no medications or procedures will be added to routine care.

Not yet recruiting

Trial opening soon.

Get Notified

Key information

Age range

18 year–85 year

Sex eligibility

All sexes

Study type

Observational

Primary location

Faculty of Medicine, Khon Kaen University

Khon Kaen, 40002, Thailand

About this study

Background: Inadvertent perioperative hypothermia (IPH), defined as a core body temperature < 36.0°C, affects 20-90% of patients undergoing general anesthesia and is associated with coagulopathy, surgical site infection, morbid cardiac events, and prolonged drug metabolism. Accurate intraoperative temperature monitoring is mandated by professional society guidelines.

Study design: Prospective, single-center, triple-method observational agreement study at Srinagarind Hospital, Khon Kaen University, Thailand. All three thermometry modalities are applied simultaneously in each participant. No intervention is assigned; temperature monitoring is additional to standard perioperative care.

Methods: After induction of general anesthesia and tracheal intubation, the nasopharyngeal (NP) probe (reference standard), TaiDoc TD-1261 infrared tympanic thermometer (test method 1), and 3M SpotOn™ zero-heat-flux forehead sensor (test method 2) are placed. Simultaneous temperature readings are recorded at 15-minute intervals from 15 minutes after induction until surgery end. For the TD-1261, two consecutive readings are taken at each time point and averaged; pinna retraction and fresh disposable probe cover are used at every measurement. The SpotOn™ sensor requires a 10-minute equilibration period before the first recorded value.

Statistical analysis: Primary analysis uses repeated-measures Bland-Altman analysis (Myles & Cui, 2007 framework) to estimate bias and limits of agreement (LoA) for each method pair against the NP probe. Pre-specified clinical acceptability criterion: LoA within ±0.5°C (Sessler, 2008). Variance of differences is decomposed into between-subject (σ²_B) and within-subject (σ²_W) components. Secondary analyses: proportional bias assessment, sensitivity/specificity for hypothermia detection (NP < 36.0°C), intraclass correlation coefficient (ICC, two-way mixed-effects model), and multivariable linear mixed-effects regression for clinical covariates. Sample size: 92 patients (Lu et al., 2016 method; 80% power; α = 0.05; SD of differences = 0.30°C for NP vs. TM; LoA acceptability threshold ±0.50°C; 15% dropout allowance).

This study provides the first institution-specific, population-specific evidence on temperature monitoring agreement in Thai surgical patients and constitutes the first published intraoperative validation of the TaiDoc TD-1261 infrared tympanic thermometer.

Who can participate

Healthy volunteers accepted: No

Only the study team can determine whether someone qualifies for participation.

Inclusion criteria

  • Age ≥ 18 years
  • We will include 3 independent groups of patients receiving general anesthesia with tracheal intubation:

Group 1: Cardiac surgery Group 2: Thoracic surgery Group 3: General surgery, including neurosurgery

  • Expected surgical duration ≥ 60 minutes
  • ASA Physical Status Classification I- III
  • Able to provide written informed consent

Exclusion criteria

  • Prior nasal surgery, nasal polyps, deviated nasal septum, or significant epistaxis history precluding safe nasopharyngeal probe insertion
  • Otitis media, tympanic membrane perforation, significant cerumen impaction, or external auditory canal deformity that would prevent reliable infrared tympanic temperature measurement
  • Facial or forehead skin lesions, burns, or known adhesive hypersensitivity at the planned 3M SpotOn™ application site, craniotomy, and head and neck surgery.
  • Baseline core temperature < 35.0°C or > 38.5°C at preoperative assessment
  • Planned use of neuraxial anesthesia as the primary or adjunct technique
  • Emergency procedures, pregnancy, or procedures involving deliberate hypothermia or cardiopulmonary bypass.

Treatment and study plan

Primary outcomes

  1. Limits of Agreement Between Each Test Method and the Nasopharyngeal Probe

    Time frame: From 15 minutes after anesthesia induction until end of surgery; measurements recorded at 15-minute intervals (estimated total observation window: 60-195 minutes per participant)

    The 95% limits of agreement (LoA = Bias ± 1.96 × SD_total) between (i) the TaiDoc TD-1261 infrared tympanic thermometer and the nasopharyngeal probe, and (ii) the 3M SpotOn™ sensor and the nasopharyngeal probe, estimated by repeated-measures Bland-Altman analysis. Pre-specified clinical acceptability criterion: LoA must be contained within ±0.5°C with both 95% CI limits of the upper and lower LoA falling within this range.

Secondary outcomes

  1. Systematic Bias (Mean Difference) of Each Test Method vs. NP Probe

    Time frame: From 15 minutes after anesthesia induction until surgery end (15-minute intervals)

    Mean difference (°C) with 95% CI between each test method and the NP probe across all paired observations, estimated from the repeated-measures Bland-Altman model

  2. Proportional Bias Assessment

    Time frame: From 15 minutes after anesthesia induction until surgery end

    Linear regression of the difference (test - reference) on the mean of both methods, with cluster-robust standard errors; Pitman's test for correlation between sum and difference

  3. Sensitivity and Specificity for Hypothermia Detection

    Time frame: From 15 minutes after anesthesia induction until surgery end

    Sensitivity, specificity, PPV, and NPV of each test method for detecting intraoperative hypothermia defined as NP temperature < 36.0°C, with AUROC estimated by DeLong's method

  4. Intraclass Correlation Coefficient

    Time frame: From 15 minutes after anesthesia induction until surgery end

    ICC (two-way mixed-effects model, absolute agreement) with 95% CI for each method pair

  5. Percentage of Paired Measurements Within ±0.5°C

    Time frame: From 15 minutes after anesthesia induction until surgery end

    Percentage of all paired simultaneous readings in which the test method falls within ±0.5°C of the simultaneous NP reference reading

  6. Clinical Covariate Association With Method Difference

    Time frame: Baseline characteristics collected before surgery; temperature data from 15 minutes post-induction until surgery end

    Association between clinical covariates (age, sex, BMI, ASA status, surgical duration, anesthetic technique, active warming use, ambient OR temperature, baseline core temperature) and the absolute method difference, estimated by multivariable linear mixed-effects regression

  7. Incidence of Device-Related Adverse Events

    Time frame: From device placement until surgery end

    Rate of epistaxis (NP probe), aural discomfort or canal injury (TD-1261), and adhesive skin reaction (SpotOn™) during the monitoring period

Study contacts

Contact information is provided by the study sponsor or research team.

Sirirat Tribuddharat, MD, PhD

CONTACT

[email protected]

66816205920

Thepakorn Sathitkarnmanee, MD

CONTACT

[email protected]

66819547622

Sponsors and collaborators

Lead sponsor

Khon Kaen University

Other

Registry information

Official study title

A Prospective Observational Method-Comparison Study of the Agreement Among Nasopharyngeal Temperature Probe, TaiDoc TD-1261 Infrared Tympanic Thermometer, and 3M SpotOn™ Zero-Heat-Flux Skin Sensor for Intraoperative Core Temperature Monitoring in Adult Patients Undergoing Elective Surgery Under General Anesthesia

Acronym: TEMP-AGREE

Important dates

Study start
2026
Primary completion
2027
Study completion
2027
First posted
Aug 6, 2026
Registry last updated