The verification of an investigational pulse oximetry board to verify pulse rate and saturation accuracy over a specified saturation range in diverse populations.
Looking for future studies?
Notify Me18 year–39 year
All sexes
Observational
Denver, Colorado, 80218, United States
To verify the SpO2 and pulse rate accuracy in comparison to a reference standard blood measurements. To achieve paired observations of SpO2 and SaO2 values over the specified SpO2 accuracy range of the prototype pulse oximeter on a group of healthy adult volunteers. The fraction of inspired oxygen (FiO2) delivered to test subjects is varied to achieve a series of targeted steadystate saturation periods. In this study, the investigational pulse oximetry PCBA system saturation measurements will be compared to saturation measurements made by a multi-wavelength CO-oximeter, taken from arterial blood samples from a diverse pool of healthy human subjects.
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
Time frame: 2.5 hours
Verification of Saturation Accuracy Saturation accuracy in a diverse subject population over a specified saturation range.
The acceptance criteria for SpO2 during low saturation conditions across the saturation range of 60-80%:
SpO2 Max-A ±3% (ARMS) Max-N ±3% (ARMS) OxySoft-N ±3% (ARMS) Max-Fast ±3%(ARMS)
Time frame: 2.5 hours
Verification of Saturation Accuracy Saturation accuracy in a diverse subject population over a specified saturation range. The acceptance criteria for SpO2 during saturation conditions across the saturation range of 70-100%: SpO2 Max-A ±2% (ARMS) Max-N ±2% (ARMS) OxySoft-N ±2% (ARMS) Max-Fast ±2%(ARMS)
Time frame: 2.5 hours
Pulse rate accuracy Max-A <=3 BPM (ARMS) Max-N <=3 BPM (ARMS) OxySoft-N <=3 BPM (ARMS) Max-Fast <=3 BPM(ARMS)
Looking for future studies?
Notify MeMedtronic - MITG
Industry
Low Saturation Verification of Investigational Pulse Oximetry Devices in Adult Volunteers
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT06264193
Behavior, Hypoxia
View Trial DetailsNCT06912230
Altitude, Appetite
Ottawa, Ontario, Canada
View Trial DetailsNCT03464695
COPD Exacerbation, COPD Exacerbation Acute
Hellerup, Denmark
View Trial DetailsNCT04055077
Hypoxia, Pars Plana Vitrectomy
Zagreb, Croatia
View Trial Details