Masimo Corporation
Irvine, California, 92618, United States
NCT Number: NCT06120777
Assess the accuracy of INVSENSOR00069's peripheral oxygen saturation (SpO2) against contemporaneous measurement from arterial blood gas (ABG) analysis.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Interventional
Not applicable
Irvine, California, 92618, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Influenced by study design:
Exclusion criteria
Influenced by device warning labels:
Influenced by study design/environment:
Noninvasive pulse oximeter
Time frame: 1-3 hours
The SpO2 accuracy of Masimo INVSENSOR00069 was determined by calculating the accuracy root mean-squared (Arms) difference between the measured values (SpO2) to the reference values (SaO2) in accordance with ISO-80601-2-61.
Arms= √(∑(i=1 to n) ((SpO2_i-SaO2_i )^2 ))/n.
Masimo Corporation
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT07365865
Deficient Keratinized Mucosa, Healthy
Cairo, Egypt
View Trial DetailsNCT04849923
Healthy
Winterthur, Canton of Zurich, Switzerland
View Trial DetailsNCT06744517
Behavior, Healthy
Toronto, Ontario, Canada
View Trial DetailsNCT05166096
Healthy
Austin, Texas, United States
View Trial Details