Masimo Corporation
Irvine, California, 92618, United States
NCT Number: NCT05674136
This study is designed to compare the equivalence of the Masimo INVSENSOR00061 to RD SET Neo for SpO2 and performance of PR and RRp as the subjects undergo a controlled desaturation protocol. This type of study design is required to evaluate the performance of pulse oximeters based upon the ISO 80601-2-61 standard which is also adopted and referenced by the FDA.
Looking for future studies?
Notify Me18 year–50 year
All sexes
Interventional
Not applicable
Irvine, California, 92618, United States
Healthy volunteers accepted: Yes
Only the study team can determine whether someone qualifies for participation.
Inclusion criteria
Exclusion criteria
(*)Physician's discretion.
Noninvasive pulse oximeter sensor
Time frame: 1-3 hours
SpO2 accuracy was determined by calculating the root mean-squared (rms) difference between the measured values (SpO2i) to the reference values (SpO2ref_i) in accordance with ISO-80601-2-61. Arms= √(∑(i=1 to n) ((SpO2_i-SpO2ref_i )^2 ))/n
Masimo Corporation
Industry
OpenTrials presents study information sourced from ClinicalTrials.gov. The official registry record should be consulted for the latest information.
View the official ClinicalTrials.gov record (opens in a new tab)This listing is for discovery and informational purposes only. It is not medical advice, does not guarantee that a study is recruiting, and does not determine eligibility. Contact the study team and a qualified healthcare professional when considering participation.
Published trials that share one or more normalized conditions with this study.
NCT05924009
Healthy
Pittsburgh, Pennsylvania, United States
View Trial DetailsNCT07374406
Healthy
Jeonju, South Korea
View Trial DetailsNCT07365865
Deficient Keratinized Mucosa, Healthy
Cairo, Egypt
View Trial DetailsNCT04849923
Healthy
Winterthur, Canton of Zurich, Switzerland
View Trial Details